Operating parameters for Agilent MP-AES 4200 system.

Element Wavelength λ (nm) Nebulizer Flow (L/min) Air Injection Flow Rate Optimal dilution factor of wine Linear domain for r2 ≥ 0.995
Zn 213.857 0.45 medium 1/10 10–200 ppb
Fe 371.993 0.65 high 1/10 3–500 ppb
Cu 324.754 0.7 high 1/10 3–2000 ppb
Ni 352.454 0.7 high 1/10 5–2000 ppb
Pb 405.781 0.75 high 1/10 25–2000 ppb
Mn 403.076 0.9 high 1/10 3–500 ppb
Mg 285.213 0.9 medium 1/100 5–2000 ppb
Ca 396.847 0.6 high 1/100 25–300 ppb
Na 589.592 0.95 high 1/100 0.25–5 ppm
K 769.897 0.75 high 1/100 0.1–5 ppm